F-Factor systematics Correction#

Input Data#

Flat-field and NSB pedestal events (usually taken with a 1 kHz frequency) from an intensity scan obtained changing the filters in front of the Calibox laser. Each run corresponds to a different flat-field intensity. These runs are tagged with run_type PEDCALIB in the run summary file. By default, a statistics of 10,000 events (per trigger type) is required in order to collect enough statistics per pixel.

How To#

F-factor systematics corrections are obtained with two steps:

1. All PEDCALIB runs from the filter scan are reconstructed as normal calibration runs with the tools used for the pixel calibration (in particular lstcam_calib_onsite_create_calibration_files_with_batch)

2. The fit of the signal intensity per gain and per pixel is performed with the tool lstcam_calib_onsite_create_fit_intensity_scan_file, which can be run with the simple command:

lstcam_calib_onsite_create_fit_intensity_scan_file -d [date] –config [config_file]

where :

It is possible to write data in a not official data-tree with the option -b [data-tree-root].

Output data#

Coefficients are written in a fit, fits.gz or hdf5 format depending on the option --output-format.

The data model and format are described below :